Pointer Model Misaligned When Using NDI Scanner and Pointer #5

Open
ben-metzger-z opened this issue Apr 25, 2024 · 0 comments
Open

Comments

@ben-metzger-z
Copy link

Hi, thanks for creating this module!
I am currently trying to perform the ASTM Phantom test on a fusionTrack500 and an NDI Polaris Vega for comparison. Everything works perfectly for the fusionTrack500, but for the Vega the initial calibration of three points often fails to acquire the second and third divots, only acquiring inconsistently with the pointer at strange angles.

If the initial calibration is completed, the onscreen model for the pointer seems to have its tip placed where the NDI pointers primary tracking sphere actually is. This makes the rest of the test impossible to perform.

Not sure if I'm missing something in the PlusLib config file or the pointer file that I need to change to set a pointer tip offset, or some way of pivoting in Slicer to get the pointers tip.