Medtrum: Priming screen returns to Change Patch screen #3365

Open
dar-kod opened this issue Jun 13, 2024 · 5 comments
Open

Medtrum: Priming screen returns to Change Patch screen #3365

dar-kod opened this issue Jun 13, 2024 · 5 comments

Comments

@dar-kod
Copy link

dar-kod commented Jun 13, 2024

I had a similar problem with several patches. This time (before 16:23 today), when I clicked to start priming, the screen immediately returned to the start screen if nothing had been started, but I could hear clicks. I could see a very short failure message, but it did not appear on the overview screen. When I tried to detect "activation in progress", the process of reading the activation status never finished. I found this very annoying. I had to discard the patch.

The first time I discarded the patch I could hear clicks, but not like before, there was a pause after every 4, 5 or 6 clicks. It finished priming and I had to activate the patch (apply the needle). I simply removed a sticker where the needle is, but then it happened again in a similar way to the first priming.

The second time I discarded the patch I managed to activate it. There was still a problem with the priming. The third priming time I wanted to be sure that the base was well connected, and it was not, as the priming did not work. The needle seems to play an important role here. I pressed the needle to get it out of the patch and a lot of insulin came out, but the priming was successful. I then applied the patch. I follow the strict instructions and wonder why there was a problem with priming the first time. I suspect I let too much air out of the patch and it couldn't start the priming process. The log file is attached.

AndroidAPS.2024-06-13_00-00-00.285.zip

@dar-kod dar-kod changed the title Medtrum: Priming screen automatically returns to Change Patch screen Medtrum: Priming screen returns to Change Patch screen Jun 13, 2024
@jbr7rr
Copy link

jbr7rr commented Jun 16, 2024

Logs are not complete, there are no logs related to any priming or activation, do you still happen to have the complete logs ?

@dar-kod
Copy link
Author

dar-kod commented Jun 16, 2024

No. For the next time, is the AndroidARS.log file that contains this information?

@jbr7rr
Copy link

jbr7rr commented Jun 17, 2024

No. For the next time, is the AndroidARS.log file that contains this information?

Everything in the log folder, best to zip the whole folder, or use the export log function in aaps itself and mail it to yourself and share that zip

@dar-kod
Copy link
Author

dar-kod commented Jun 24, 2024

It happened again, very similar. What is new, and what I didn't report before but have seen, is that after the priming screen returned to the attached screen (which comes after the first Change Patch screen), the message "Pump time updated" came up at the same time. Please see a screenshot of this. I didn't apply the patch immediately, but waited perhaps 3–5 minutes to prepare to apply it. It seems that the driver forgets the patch because of the waiting.

1719234780257

I still managed to activate the patch after pressing discard and repeating the process.

The full zipped log files are attached.

files.zip

@dar-kod
Copy link
Author

dar-kod commented Jul 18, 2024

Sorry. My zip file from last time was empty.

I had a similar problem again. This time I had the patch on the body when the screen that comes after the Priming screen went back to the Change Patch screen. This happened at about 8:05 in the morning. Again, the process of reading the activation status never finished. I tried to see if Exit would help, but it didn't. I tried detach the pump base 2 times. If I remember correctly, the first time the pump state showed PATCH_FAULT2 (8:16) after unsuccessful priming, the priming was not completed after reaching 100% (8:12). I understood that there was some misunderstanding between the driver and the patch, but I didn't want to lose the patch. When I tried to read the activation status, the only option was to deactivate the patch. So I understood that I had to insert the needle to try the last option. I did so and exited AAPS. The patch was ready to be activated again with the previously removed pump base (I don't remember exactly when I did this), but with a needle in the body. I simply added 2 units without bolusing.

files.zip